Shielded dispensing isolator with front manipulators and a side work shelf

Dispensing isolator

A shielded Class A isolator for filling, calibrating and fractioning radiopharmaceuticals into vials and syringes.

Air quality
Class A
Vials per batch
10–15
Shielding
75 or 100 mm Pb

Where it fits in production

  1. Isotope production
  2. Synthesis
  3. DispensingThis isolator
  4. Quality control
  5. Radiation protection

Overview

The dispensing stage of a GMP radiopharmacy line. Laminar flow covers the entire main chamber, and the isolator can house automatic dispensing modules or fractioning systems in sterile conditions.

  • Operator protection

    Complete shielding, an air seal between the chambers and high-sensitivity tele-pliers protect the operator from ionising radiation.

  • Sterile material transfer

    A ventilated pre-chamber brings materials in without breaking the sterile environment of the main chamber.

  • Dose measured in the cell

    A built-in dose calibrator with its own touch-screen console measures each dose before it leaves the isolator.

  • Tested and traceable

    Connections for filter leak testing, bubble-point testing and hydrogen-peroxide decontamination support routine cGMP checks.

Specifications

Typical configuration. The final specification is set for your project.

Model
BBST
Function
Filling, calibrating and fractioning radiopharmaceuticals in vials and syringes
Air quality
Class A, with laminar flow across the whole main chamber; cGMP compliant
Typical batch
10–15 vials; on request, 3–5 syringes of 5 ml
Configuration
Single, double, triple or quadruple chamber
Shielding
75 mm or 100 mm lead
Dose calibrator
2 Ci or 20 Ci versions, with touch-screen console
Handling
High-sensitivity tele-pliers on ball bearings
Fluid lines
AISI 316L stainless steel with ball shut-off valves; airtight connections for radioactive fluids
Pressure control
Automatic internal pressure adjustment by modulating valves
Testing
Filter leak (DOP) test connections; semi-automatic bubble-point test
Decontamination
VPHP connections to the main chamber and pre-chambers
Product extraction
Drawing system for vials and syringes
Control
7-inch touch-screen operator panel

Solutions that use it

All solutions

Planning a GMP radiopharmacy?

Tell us where your project stands, from a first idea to an existing facility that needs equipment or support.